15 Specification  
Measuring range 
air pressure
0 ... 11000 mbar abs.,  
resolution 1 mbar  
0 ... 11.000 bar abs.,  
resolution 0.001 bar 
0 ... 8250 mmHg abs., 
resolution 1 mmHg 
0 ... 160.00 PSI abs., 
resolution 0.02 PSI 
Measuring media: 
non corrosive gases 
Max. overload: 
13 bar abs. 
Accuracy: (±1 digit) 
+/- 3 mbar or +/- 0.10 % typ., whichever is higher (at nominal 
temperature 25°C) 
+/- 0.30 % FS temperature influence 0 to 50°C 
Pressure connection: 
connecting nozzle made of nickel-plated brass on top of the device,  
for tubes 6x1mm (4mm innerØ), approx. 11mm projecting 
Measuring frequency: 
1 measurement per second 
Offset and scale: 
digital offset and scale adjustment for air pressure measurement  
Min-/max-value memory:  min- and max- value are stored 
Sea level correction: 
The displayed value can be referenced to sea level. Therefore the 
current height above sea level has to be entered. 
Zero-function: 
relative measurement: the displayed value is set to “zero”  
Display: 
approx. 13  mm high, 4½-digit LCD display 
Pushbuttons: 
3 membrane keys for on/off, min-/max-value request, offset adjustment   
Working conditions: 
-25 to 50 °C; 0 to 80 % RH (non condensing) 
Storage temperature: 
-25 to 70 °C 
Power supply: 
9V-battery, type 6F22 (included in scope of delivery) 
    Current consumption: 
ca. 60 µA  (Operating time with standard zinc-carbon battery >7500 
hours 
    Display for weak battery:  automatically if battery consumed "BAT", warning: "LOBAT" 
Auto-off function: 
If activated: device will be automatically switched off if not operated for 
longer time (selectable 1..120min)  
Housing: 
impact resistant ABS housing, front side IP65 
Dimensions: 
approx. 106 x 67 x 30 mm (H x W x D) , without connecting nozzle  
Weight: 
approx. 135 g incl. battery 
EMC: 
The device corresponds to the essential protection ratings established 
in the Directives of the European Parliament and the council on the 
approximation of the laws of the member states relating to the 
electromagnetic compatibility (2004/108/EG). Additional error: <1% 
16 Disposal notes 
 
Dispense exhausted batteries at destined gathering places. 
This device 
must not be disposed as “residual waste”. 
To dispose this device, please send it directly to us (adequately stamped). 