Cleaning Recommendations 
When cleaning the probe, take care not to scratch or damage the sensing surface or the 
platinized electrode surfaces. 
 
Contaminant Cleaning 
Solution 
Instructions 
Water soluble substances 
Deionized water 
Soak or scrub with a soft 
brush. Recondition in 4 or 7 
buffer for 1 hour. 
Grease & Oil 
Warm water and household 
detergent 
Soak or scrub with a soft 
brush, maximum of 10 
minutes. Rinse thoroughly 
with DI water, recondition in 
4 or 7 buffer for 1 hour. 
Heavy grease & Oil 
Alcohol 
Maximum of 5 minute soak, 
scrub with a soft brush. 
Rinse thoroughly with DI 
water, recondition in 4 or 7 
buffer for 1 hour. 
Lime and hydroxide 
coatings 
10% acetic acid 
Soak until coating dissolved, 
maximum of 5 minutes. . 
Rinse thoroughly with DI 
water, recondition in 4 or 7 
buffer for 1 hour. 
Please Note: Since the EC500 does not have a refillable reference electrolyte chamber, it is 
important not to soak the electrode in the above solutions for more than the recommended 
times. To do so may cause a reference potential shift, which will cause degradation in 
performance or failure.