Gw Instek GDS-2074A 4-channel oscilloscope, Digital Storage oscilloscope, GDS-2074A Data Sheet

Product codes
GDS-2074A
Page of 261
 
 
GDS-2000A Series User Manual 
242
 
How to use SPC function 
 
Background 
Signal Path Compensation (SPC) is used to 
compensate the internal signal path due to 
ambient temperature. SPC is able to optimize the 
accuracy of the oscilloscope with respect to the 
ambient temperature. 
Panel Operation 
1.  Press the Utility key. 
Utility
 
2.  Press System from the bottom 
menu.  
 
 
3.  Press SPC from the side menu. A 
message showing a brief 
introduction to SPC appears on the 
screen. 
 
Note 
Disconnect all probes and cables from all channels 
before calibrating. 
The DSO needs to be warmed up for at least 30 
minutes before using the SPC function. 
 
4.  Press Start on the side menu to 
start SPC calibration. 
 
 
5.  The SPC Calibration will proceed one channel 
at a time, from channel 1 to channel 4.