Gw Instek GDS-2074A 4-channel oscilloscope, Digital Storage oscilloscope, GDS-2074A Data Sheet

Product codes
GDS-2074A
Page of 261
 
 
MAINTENANCE 
243 
Vertical Accuracy Calibration 
 
Panel Operation 
1.  Press the Utility key. 
Utility
 
2.  Press System from the bottom 
menu.  
 
 
3.  Press more 1 of 2 from the side 
menu. 
 
 
4.  Press Self Cal on the side menu. 
 
 
5.  Press Vertical on the side menu. 
 
 
6.  A message appears to “Now performing 
vertical calibration…Set CAL to the channel, 
then press the Vertical key”. 
 
7.  Connect the calibration signal from the rear 
panel to the Channel 1 input with a BNC cable. 
 
 
CH1
 
 
8.  Press Vertical again after 
connecting CAL to the channel 1 
input.