Voltcraft CPPS-160-42, 1 Output Variable DC Power Supply, , Bench CPPS-160-42 User Manual

Product codes
CPPS-160-42
Page of 143
59
15. SENSE FUNCTION
The sense function is an automatic voltage control for the connection sockets MAIN OUTPUT (6).  
In addition two separate measurement lines are connected parallel to the connection lines. The 
voltage drop, which occurs at the connection lines, is measured on both these measurement 
lines. This voltage drop automatically balances the laboratory power supply so that the voltage 
actually set fits the consumer.
1.  Switch off the power supply and consumer.
2.  Always first connect the power supply lines of the power supply to the consumer. Please
note the correct polarity.
3.  Press the terminal release inside with a small screw driver on the rear connection sockets 
Remote Sensing (10) and stick the cables into the clamp openings. Check the secure 
position.
4.  Now connect both sense lines to the consumer at the correct polarity. The conductor cross 
section for the sense lines must always amount to at least 0.34 mm
2.
5.  Always loosen the connections in reverse order (first the sense lines and then the
connection lines).
R
emot
e
Sensing
Please ensure that the sense cables are connected as near as possible to 
the connection point of the consumer. Please note the correct polarity.
Never short circuit the sense lines.