Gw Instek LCR-816 Digital-Multimeter, DMM, LCR-816 Data Sheet

Product codes
LCR-816
Page of 91
  
BASIC 
MEASUREMENT 
31 
Parameter Configuration 
Measurement Speed 
 
Measurement 
Speed  
The LCR-800 series support 3 different 
measurement speeds: slow, medium or fast at 
approximately 1, 5 or 12 (LCR-829/827/826) 
measurements per second. The faster the 
measurement speed, the lower the accuracy. 
Conversely the slower the measurement speed, the 
higher the accuracy.  The measurement speed and 
accuracy are dependent on the mode, voltage and 
frequency. For detailed information, see the 
specification table on page 136. 
LCR-817/ 819/ 821 
Accuracy 
Measurements/second 
 
Slow 
0.05%  
At least 1 
 
Medium 
0.1% 
At least 3 
 
Fast 
0.24% 
At least 7 
LCR-816/826/827/829 Accuracy  Measurements/second 
 
Slow 
0.1%  
At least 1 
 
Medium 
0.2% 
At least 3 
 
Fast 
0.48% 
At least 7 
Panel operation 
1.  From the main menu, press the 
SPEED menu key to cycle between 
the various speeds. 
 
 
 
 
 
   LCR-800 
User 
Manual 
32
 
Displayed measurement unit 
 
Measurement 
units  
All measurement unit results can be displayed as 
the absolute values, delta values or delta 
percentage values. 
Value will show the absolute value of the 
measurement in Ohms (Ω), Henries (H) or Farads 
(F).  The primary measurement has resolution of 5 
digits; the secondary has a resolution of 4 digits (θ, 
2 digits).  
Delta% will show the percentage deviation of L, C, 
R or Z from a nominal (stored) value. 
Delta will show the deviation from a nominal 
value as an absolute value in Ohms (Ω), Henries 
(H) or Farads (F). 
  
Units 
 Value 
Ω, H, F 
 Delta 
Absolute deviation (Ω, H, F ) 
 Delta% 
deviation 
Panel operation 
1.  From the main menu, press the 
DISPLAY menu key to cycle 
between the display types.