BASIC 
MEASUREMENT 
33 
Measurement Modes  
 
Measurement 
mode  
The LCR-800 has a number of different 
measurement modes. Primary and secondary 
measurements are displayed on the screen 
simultaneously. For detailed information 
regarding the measurement combinations, see the 
specifications on page 136. The measurement 
combinations are shown in the table below. 
 (C/D) 
Capacitance/Dissipation 
 (C/R) 
Capacitance/Resistance 
 
 (L/R)* 
Inductance/Resistance 
 (L/Q) 
Inductance/Quality 
factor 
 (Z/θ)* Impedance/Angle 
 (R/Q) 
Resistance/Quality 
factor 
Panel operation 
1.  From the main menu, press the 
MODE menu key to cycle between 
the different modes. 
 
Note 
*Only the LCR-821 can select L/R and Z/θ 
measurement modes.  
Select Equivalent Circuit Type 
 
Background 
Series or Parallel equivalent circuits can be 
selected.  Not all measurement modes can be used 
with both series and parallel equivalent circuits. 
For details about circuit types see the circuit theory 
chapter on page 124. 
 Measurement 
type 
Series Parallel 
 Capacitance 
(C) 
 
 
 Inductance 
(L) 
 
 
 Impedance 
(Z) 
 
 
 Resistance 
(R) 
 
 
   LCR-800 
User 
Manual 
34
 
Panel operation 
1.  From the main menu, press the 
CIRCUIT menu key to cycle 
between the series or parallel 
equivalent circuits. 
 
 
 
 
 Series 
 
 
 
Parallel 
 
 
Set Bias voltage 
 
Background 
Voltage bias can be set internally or externally. An 
internal voltage bias of 2 volts is normally applied 
to a DUT. External voltage bias is able to accept 0 
to 30 volts with a maximum current of 200mA. For 
external bias voltage connections see page 22. 
When measuring a DUT, please allow 1 second to 
stabilize a DUT after a bias voltage is applied. In 
general a bias voltage should only be applied to 
capacitors. If a bias voltage is applied to devices 
with low impedance, inaccurate measurements 
will occur.   
Note 
When an external voltage is applied, constant voltage 
mode (C.V.ON) must be enabled, page 39.