PROGRAMMING 
91 
 
SPEED  
Command/Query 
 
The speed command sets the measurement speed of the instrument. 
The faster the measurement speed the lower the accuracy. This 
command also queries the current measurement speed. 
Syntax MAIN:SPEE:<string><^END^M>or<^J^M> 
 
Parameter 
<string> Speed 
SLOW Slow 
MEDI Medium 
FAST Fast 
Example MAIN:SPEE:SLOW<^END^M> 
 
Set the measurement speed to slow. 
Query Syntax 
MAIN:SPEE?<^END^M>or<^J^M> 
Return String 
<string> Speed 
MAIN:SPEE:SLOW<^END> 
Slow 
MAIN:SPEE:MEDI<^END> 
Medium 
MAIN:SPEE:FAST<^END> 
Fast 
Query Example 
MAIN:SPEE?<^END^M> 
MAIN:SPEE:MEDI<^END>  
 
Medium measuring speed is returned. 
DISPLAY  
Command/Query 
 
The display command sets the displayed measurement as a value or 
as an offset from a nominal value (Delta or Delta%) 
Syntax MAIN:DISP:<string><^END^M>or<^J^M> 
Parameter 
   LCR-800 
User 
Manual 
92
 
<string> Display 
VALU Unit 
Value 
DELP Delta 
DELT Delta 
 
Example MAIN:DISP:VALU<^END^M> 
 
Set the display to Value 
Query Syntax 
MAIN:DISP?<^END^M>or<^J^M> 
Return String 
<string> Display 
MAIN:DISP:VALU<^END> Value 
MAIN:DISP:DELP<^END> Delta 
MAIN:DISP:DELT<^END> Delta 
 
Query Example 
MAIN:DISP?<^END^M> 
MAIN:DISP:VALU<^END> 
 
Currently the display is set at value. 
MODE 
Command/Query 
 
The mode command sets the measurement mode of the LCR-800.  
Syntax MAIN:MODE:<string><^END^M>or<^J^M> 
Parameter 
<string> Primary 
Measurement Secondary 
Measurement 
RQ Resistance  Quality 
factor 
CD Capacitance Dissipation 
factor 
CR Capacitance Resistance 
LQ Inductance  Quality 
factor 
LR* Inductance  Resistance 
ZQ* Impedance  Angle 
*For the LCR-821 only