5 Performing the measurement 
19 
5.4
  Image type 
 
This function is only available for the testo 870-2 version. 
 
The display can be switched between infrared image and real image (digital 
camera). 
- Select 
Image type
 function. For how to proceed: see Getting to know 
 
The image type displayed switches alternately between the infrared 
image view and real image view. 
 
 
5.5
  Setting the scale 
 
Manual scaling can be activated instead of automatic scaling (continuous 
automatic adjustment to the current min./max. values). The scale limits can be 
set within the measurement range.  
The activated mode is displayed bottom right:   automatic scaling,   manual 
scaling. 
 
 
Auto-scaling continuously adjusts the scale to the readings presented, 
and the colour assigned to a temperature value changes.  
In manual scaling fixed limit values are defined, and the colour 
assigned to a temperature value is fixed (important for visual image 
comparisons).  
The scaling affects how the infrared image appears on the display, but 
has no effect on the recorded readings. 
 
 
Setting automatic scaling 
- Select 
Scale
 function. For how to proceed: see Getting to know the 
menu, page 11. 
 
- Move the 
Joystick
 to the left until   
Auto
 is selected and then press 
OK
 
Automatic scaling is activated.    is displayed bottom right. 
 
 
Setting manual scaling 
The lower limit value, the temperature range (upper and lower limit value 
simultaneously) and the upper limit value can be set. 
- Select 
Scale
 function. For how to proceed see Getting to know the 
menu, page 11.