5 Performing the measurement 
20 
 
- Move the 
Joystick
 to the left/right until   (lower limit value) is 
selected. 
- Move the 
Joystick
 up/down to set the value. 
 
2.1 
- Move the 
Joystick
 to the left/right until   (lower limit value) and   
(upper limit value) are selected. 
- Move the 
Joystick
 up/down to set the values. 
 
2.2 
- Move the 
Joystick
 to the right until   (upper limit value) is selected. 
- Move the 
Joystick
 up/down to set the value. 
 
- Press 
OK
 
 
Manual scaling is activated.   is displayed bottom right. 
 
 
5.6
  Setting emissivity and reflected 
temperature 
 
testo  870-2: this function is only available if the 
Image type
 is set to infrared 
image. 
 
You can choose between user-defined emissivity and 8 materials with 
permanently set emissivity. The reflected temperature (RTC) can be set 
individually. 
 
Other materials can be imported into the instrument from an existing 
list using the PC software. 
 
 
Emissivity information: 
The emissivity describes the capability of a body to emit electromagnetic 
radiation. This is material-specific and must be adapted for correct 
measurement results. 
Non-metals (paper, ceramic, gypsum, wood, paints and coatings), plastics and 
food have high emissivity, which means that the surface temperature can be 
easily measured using infrared. 
Because of their low or non-uniform emissivity, bright metals and metal oxides 
only have limited suitability for infrared measurement. Highly inaccurate 
measurements should be expected. A remedy for this is coatings that increase 
emissivity, e.g. paint or emission adhesive (accessory: 0554 0051), which must 
be applied to the object to be measured.