Lecroy WaveJet 354A 500 MHz 4 Channel Oscilloscope WaveJet 354A Data Sheet

Product codes
WaveJet 354A
Page of 67
WaveJet 300A Series  
52
 
 
WJ-A-GS-E Rev A
 
PARAMETER MEASUREMENTS 
Waveform analysis typically begins with the measurement of parameters. Parameter 
measurement tools determine a wide range of waveform properties. Use them to 
automatically calculate many attributes of your waveform, like rise time, rms voltage, 
and peak-to-peak voltage, for example. 
You can make common measurements on one or more waveforms. Parameter values 
are displayed below the grid. 
 
Measure Modes 
Standard Vertical Parameters 
  Maximum -- Measures highest point in waveform. Unlike top, it does not 
assume the waveform has two levels. 
  Minimum -- Measures the lowest point in a waveform. Unlike base, it does not 
assume the waveform has two levels. 
  Peak-Peak -- Difference between highest and lowest points in the waveform.  
  RMS -- Root Mean Square of data between the cursors -- about the same as 
sdev for a zero-mean waveform. 
  Cycle RMS -- Cyclic root mean square: Computes the square root of the sum of 
squares of data values divided by number of points. Contrary to rms, 
calculation is performed over an integer number of cycles, eliminating bias 
caused by fractional intervals. 
  Mean -- Average of the data for a time domain waveform. Computed as 
centroid of distribution for a histogram. 
  Cycle Mean -- Computes the average of the waveform data. Contrary to mean, 
computes the average over an integer number of cycles, eliminating bias 
caused by fractional intervals. 
  Top -- Higher of two most probable states, the lower being base; it is 
characteristic of rectangular waveforms and represents the higher most 
probable state determined from the statistical distribution of data point values in 
the waveform. 
  Base -- Lower of two most probable states (higher is top). Measures lower level 
in two-level signals. Differs from min in that noise, overshoot, undershoot, and 
ringing do not affect measurement.