Lecroy WaveJet 354A 500 MHz 4 Channel Oscilloscope WaveJet 354A Data Sheet

Product codes
WaveJet 354A
Page of 67
Getting Started Manual 
WJ-A-GS-E Rev A 
 
53
 
  Top-Base -- Measures the difference between upper and lower levels in two-
level signals. Differs from pkpk in that noise, overshoot, undershoot, and ringing 
do not affect the measurement. 
  +Overshoot -- Amount of overshoot following a rising edge specified as 
percentage of amplitude. 
  -Overshoot -- Amount of overshoot following a rising edge specified as 
percentage of amplitude. 
Standard Horizontal Parameters 
  Rise Time 10-90% -- Detects the first rise to pass through 50% of the amplitude 
(top - base) of the waveform within the measurement section, and then 
measures the time of transition from 10% to 90% at that rise. 
  Rise Time 20-80% -- Detects the first rise to pass through 50% of the amplitude 
(top - base) of the waveform within the measurement section, and then 
measures the time of transition from 20% to 80% at that rise. 
  Fall Time 80-20% -- Detects the first fall to pass through 50% of the amplitude 
(top - base) of the waveform within the measurement section, and then 
measures the time of transition from 80% to 20% at that fall. 
  Fall Time 90-10% -- Detects the first fall to pass through 50% of the amplitude 
(top - base) of the waveform within the measurement section, and then 
measures the time of transition from 90% to 10% at that fall. 
  Frequency -- Period of cyclic signal measured as time between every other pair 
of 50% crossings. Starting with first transition after left cursor, the period is 
measured for each transition pair. Values then averaged and reciprocal used to 
give frequency. 
  Period -- Period of a cyclic signal measured as time between every other pair of 
50% crossings. Starting with first transition after left cursor, period is measured 
for each transition pair, with values averaged to give final result. 
  No. of +Pulses -- An integer number of positive pulses. 
  No. of -Pulses -- An integer number of negative pulses. 
  +Pulse Width -- Measures the time from the first rise until the first fall to pass 
through 50% of the amplitude (top-base) of the waveform within the 
measurement section. 
  -Pulse Width -- Measures the time from the first fall until the first rise to pass 
through 50% of the amplitude (top-base) of the waveform within the 
measurement section. 
  Duty Cycle -- Width as percentage of period.