9  Adjusting of the instrument 
  The cell correction may change due to natural ageing or deposits at the measuring cell. In case 
you have an accurate reference liquid, you may adjust the instrument by changing the cells 
scale adjustment accordingly. 
  The temperature measuring is very stable and has to be adjusted in very seldom cases only. 
  To adjust the instruments according to Your needs proceed like follows:: 
  Switch off instrument. 
  Switch on  
and press hold-key during the segment test (                  ), until the display shows the first 
paramet
er “SCL” 
  If a parameter should be edited, press key up or down ( 
 ), the setting of the parameter 
will be shown and can be changed via up/down keys and receipt the changed parameter with. 
  Jump to the next parameter with key 
 
 
Parameter 
Value 
Meaning 
key
 
 
keys
 
 
 
 
Scale correction of measuring cell 
0.800..1.200 
Adjustment of the measuring cells scale 
 
Offset of temperature 
-
2.0…2.0 °C 
-
3.6…3.6 °F 
Offset setting of temperature measurement (see below) 
 
Scale correction of temperature 
-
5.00…5.00 % 
Scale adjustment of temperature measurement 
Display = (measured value - offset) * ( 1 + scale adjustment / 100) 
Pressing 
 again stores the settings, the instruments restarts (segment test) 
Please note:  If there is no key pressed within the menu mode within 2 minutes, the 
adjusting will be cancelled, the entered settings are lost
10 System massages 
Er. 1 =   measuring range has been exceeded 
--    =    Sensor Error 
Er. 7 =   System fault - the device has detected a system fault (defective or far outside allowable 
ambient temperature range) 
If „BAT” is displayed at the left side of display, the battery is weak; measuring can be continued for 
a short period. 
If “bAt” is displayed in the main display the battery is used up and needs to be replaced. 
Measuring is no more possible. 
11 Inspection of the accuracy / Adjustment Services 
Accuracy can be inspected with test-solutions (extra equipment). 
 
If the precision is not more within the admissible range, we suggest to send the device to the 
manufacturer for a new adjustment.