Greisinger GLF 100RW Meter for reading performance ability GLF 100 RW ±(1% + 1 Count) -5 to +100 °C 600330 Data Sheet

Product codes
600330
Page of 8
H66.0.02.6C-03 
Operating Manual GLF 100 RW 
Page 8 of 8 
13 Specification 
Measuring 
 
 
  Principle 
 
Conductivity measuring via 2 pole stainless steel electrode    
  Meas. Ranges   
Range 
Resolution  
Conductivity 
0.000 ... 2.000 µS/cm 
 
001 µS/cm 
0.00 ... 20.00 µS/cm 
 
0,01 µS/cm 
0.0 ... 100.0 µS/cm  . 
 
0.1 µS/cm 
Resistivity 
0.0100...0.2000 MΩ*cm 
0.010...2.000 MΩ*cm 
0.01...20.00 MΩ*cm 
Temperature 
-5.0 ... 100.0 °C 
Electrode 
Concentric 2 pole stainless steel electrode (1.4404, 1.4435), 
integrated temperature sensor, outer diameter 12 mm, minimum 
immersion depth 25 mm 
Cable length 1.2 m. 
  Accuracy 
 
±1 Digit (at nominal-temperature) 
Conductivity:  
typ. better than ± (1,0 % + 0,5 % FS) 
Temperature  
±0.3 K 
  T.-compensation:  
linear temperature compensation with adjustable coefficient 
non-linear temperature compensation according to EN 27888  
temperature compensation of weak NaCl solutions acc. to EN 60746-3 
reference temperatures 20 °C and 25 °C 
Nominal temperature 
25 °C 
Ambient 
Temperature instrument -25 ... +50 °C 
Temperature cell -5 ... +80 °C (short time 100 °C) 
Relative humidity instrument 0 ... 95 %RH (non condensing) 
Storage temperature  
-5 ... +50 °C  
Housing 
Dimension: 110 x 67 x 30 mm (H x W x D)  
impact resistant ABS, membrane keyboard, transparent panel.  
Front side IP65 
Weight 
   
approx. 155 g 
Power Supply      
9V-Battery, type 6F22 (zinc-carbon battery included) 
 
   
Operating time with zinc-carbon battery >200 hours 
Power Consumption  
<1.5 mA 
Display  
   
ca. 11 mm high, 4½-digit LCD-display with additional segments 
Pushbuttons  
3 membrane keys for on/off switch, menu operation, min/max/ hold-
function etc. 
Hold Function      
Press key to store current value. 
Automatic-   
 
Device will be automatically switched off if not operated for longer time 
Off-Function 
 
(adjustable from1...120min) 
 
EMC:  The device corresponds to the essential protection ratings established in the Directives of 
the European Parliament and of the council on the approximation of the laws of the memberstates 
relating to the electromagnetic compatibility (2004/108/EC). EN 61326 +A1 +A2 
(Appendix B, class B), additional error: < 1 % FS.