Voltcraft MS-1300 Digital Luxmeter 0.1 - 50 000 lx MS-1300 Leaflet

Product codes
MS-1300
Page of 1
:
:
:
Calibration Procedure
1 0 calibration:
set to LUX range, cover the LUX sensor hole with 
LCD displays same reading (1000LUX) as the standard LUX meter displays.
4 20000LUX calibration:
set to 20000LUX range, following the precedure of step 2, adjust VR4 until 
set to 2000LUX range, following the precedure of step 2, adjust VR3 until 
 open the LUX sensor cover of the DT-1300 LUX meter, and put it
onto the test frame, set to 200LUX range, adjust VR1 on the PCB, 
until LCD displays same reading (100LUX) as the standard LUX meter displays.
3 2000LUX calibration:
2 200LUX calibration:
 output a certain light with a standard LUX calibrator, 
light orbit until the LCD display 100LUX, write down the reading.
put the standard LUX meter onto the test frame, and move the 
a black label to prevent any light in induced. 
LCD displays same reading (10000LUX) as the standard LUX meter displays.
Light Meter
:WI-ENG K0036
:  
:  
DT-1300
V1.0
:     
2009 7 30
 LUX
:    1 
 1 
:     
:
5 50000LUX calibration:
set to 50000LUX range, following the precedure of step 2, adjust VR5 until 
LCD displays same reading (50000LUX) as the standard LUX meter displays.
Adjust VR2 until LCD display 0. All ranges display 0.
200LUX adjustor 
VR1
50000LUX adjustor 
VR5
20000LUX adjustor 
VR4
2000LUX adjustor 
VR3
0 adjustor 
VR2