Bk Precision BK-8500 Electronic Load 1 mA - 30 A 0.1 - 120 Vdc 0 - 300 W BK-8500 Data Sheet

Product codes
BK-8500
Page of 76
Models 8520/8522/8524/8526 (2400W & 5000W)
Parameter
8520
8522
8524
8526
Input rating 
Voltage
0 – 120 V
0 – 500 V
0 – 60 V
0 – 500 V
Current 
0 – 240 A
0 – 120 A
0 – 240 A
0 – 120 A
Power
2400 W
5000 W
8520/8522/8524/8526 common characteristics
Parameter
Range
8520
8522
8524
8526
Accuracy
Resolution
CV Mode
Regulation
0.1-18 V
 ±(0.05%+0.02% FS)
1 mV
0.1 V to Vmax
 ±(0.05%+0.025% FS)
10 mV
CC Mode
Regulation
0-24 A
0-12 A
0-24 A
0-12 A
 ±(0.1%+0.1% FS)
1 mA
0 – max Current
 ±(0.2%+0.15% FS)
10 mA
Current Measurement
0-12 A
0-3 A
0-24 A 
0-12 A
 ±(0.1% + 0.1% FS)
1 mA
0 – max. Current
 ±(0.2%+0.15% FS) 
10 mA
Voltage Measurement
0 – 18 V
8522/8526: 
(0.02% + 0.02% FS) 
8520/8524:
(0.02% + 0.025% FS) 
1 mV
0 - Vmax
 ±(0.02% + 0.025%FS)
10 mV
All DC load models
common characteristics for all DC load models
Parameter
Range
Accuracy
Resolution
CR Mode Regulation
Input current ≥FS 10%
Input Voltage≥FS 10%
0.1 -10 Ω
±(1%+0.3% FS)
0.001 Ω
10-99 Ω
±(1%+0.3% FS)
0.01 Ω
100-999 Ω
±(1%+0.3% FS)
0. 1 Ω
1K-4 KΩ
±(1%+0.8% FS)
1 Ω
CW Mode Regulation
Input current ≥FS 10%
Input Voltage≥FS 10%
0-100 W
±(1%+0.1% FS)
1 mW
100 W – max Power
±(1%+0.1% FS)
100 mW
Power Measurement
Input current ≥ FS 10%
Input Voltage ≥ FS 10%
0-100 W
±(1%+0.1% FS)
1 mW
100-max Power
±(1%+0.1% FS)
100 mW
Battery testing function
Input=0.1 V – 120 V         Max measurement capacity= 999 A/H
Resolution =10 mA      Timer range=1~60000 sec
Transition Mode
                 Range of Frequency 0.1 Hz-1 kHz   Frequency error rate
<0.5%
NOTE:  Specifications and information are subject to change without notice. Please visit 
www.bkprecision.com for the most current product information.
Environmental Conditions 
This instrument is intended for indoor use in a pollution degree 2 environment.  Operating 
environmental limits are as follows: 
8500 DC Load Series                    
Version:  September 2, 2009 
Page 16 of 76