Keys
Display
current value.  Press the 1 and 0 keys to enter 10 A.
Enter
Enters the 10 A value, then displays  
WIDTH B = X.XMS
.  You're being 
prompted for the duration of the 10 A load -- press the 1 and 0 key for 10 
ms.
Enter
Enters the 10 ms duration, then displays 
:CONTINUOUS
:PULSE
, or 
:TOGGLED
. Use the 
△ or ▽ arrow keys to display 
:PULSE
.
Enter
Returns to the standard display of voltage and current.
To activate this pulse transient condition, press Shift + Tran, then press On/Off.  The load will start 
at the A level of current you entered (5 A).  Press Shift + Trigger to cause the load to switch to the 
B level of current (10 A).  The 10 A current will last for the designated 10 ms, then switch back to the 
A level (5 A).
NOTE:  This example requires the :CONFIG:TRIGGER menu item be set to 
:IMMEDIATE.  If set 
to :EXTERNAL, you'd trigger the instrument with a high TTL level signal on the trigger terminals on 
the rear panel.  If set to :BUS, you'd trigger the instrument with a software trigger.
Toggled transient operation
In toggled transient operation, the load starts at the stored parameter for the mode.  When a trigger 
is received, the load switches to the B value.  When another trigger is received, the load switches to 
the A value.  It stays at the A value until another trigger is received, at which point it switches to the 
B value.  Further triggers cause the load to switch between the A and B values.  Here's an example:
10A
5A
TRG
TRG
Toggled Transient Operation
Here are the keystrokes needed to setup this example:
Keys
Display
I-set
Standard display of voltage and current.  Enter a value of 5 A (you may 
need to press the I-set key again).  Press Enter.
Enter
Enters the CC mode's value.  The standard display is shown.
Shift + S-Tran
LEVEL A= X.XXXA
  You are being prompted for the first current value. 
Press the 5 key for 5 A.
Enter
Enters the 5 A value, then displays 
WIDTH A = X.XMS
.  You're being 
prompted for the duration of the 5 A load.  In the toggled mode of 
operation, this width is ignored, so enter any convenient value.
8500 DC Load Series                    
Version:  September 2, 2009 
Page 35 of 76