Bk Precision BK-8500 Electronic Load 1 mA - 30 A 0.1 - 120 Vdc 0 - 300 W BK-8500 Data Sheet

Product codes
BK-8500
Page of 76
Keys
Display
2.5 Enter
SHORT OFF 
 We select short on and then press Enter.
 Enter
READBACK A
  We want to readback the current, so we press Enter.
Enter
MIN 2= 4.950A 
 Our minimum value is 2.0 A.
2 Enter
MAX 2= 5.050A
  We choose 2.5 A as the maximum value.
2.5 Enter
DELAY 2= 3.0<S>
  We'll wait 2 seconds for the maximum current.
2 Enter
STORE TEST FILE1
  The 1 is underlined, meaning you can choose 
which test file number to store this test to.  You can choose any number 
between 1-8.  Here, we will use 1, so just press Enter.
Enter
EDIT TEST FILE 
 
Esc Esc
Exit the menu.  The normal display is shown (V and A).
The test file has been created and saved in file position 1.  Now we will run the test on the device.
Keys
Display
Shift + Menu
:CONFIG
 
▽ ▽
:LIST SET
Enter
:MODE SET
   
▽ ▽ ▽
:CALL TEST FILE
Enter
RECALL 1
  We're being prompted for which file number to recall.  Since 
we saved to file number 1, we just press Enter.
Enter
CALL TEST FILE
  The test file is now loaded, so we return to the normal 
display.
Esc Esc
Exit the menu.  The normal display is show (V and A).
Shift I-set
NAME:TEST FILE 1
  We've entered automatic testing mode.  The display 
is showing which test file we're using.
Shift Trigger
This starts the test.  You'll see the values sequentially displayed.  Then 
you'll either see 
PASS
 or 
FAULT
.  You can press Shift Trigger again for 
another test run or press Esc to return to the normal display.
Esc
Return to the normal display.
Triggering
Triggering is used with the transient and dynamic conditions to allow synchronization of the DC
Load'
s behavior with other events.  There are three types of triggers you may use (set in the 
CONFIG:TRIGGER SOURCE menu):
Trigger Type
Explanation
IMMEDIATE
An immediate trigger is created by pressing Shift + Trigger on the 
8500 DC Load Series                    
Version:  September 2, 2009 
Page 39 of 76