Trigger Type
Explanation
front panel keys.
EXTERNAL
An external trigger is a TTL high signal applied to the Trigger 
connection on the back panel.  This TTL signal must last for more than 
5 ms. A trigger applied to this input can be used to change settings 
(voltage, current , resistance), toggle between settings in transient-
toggle mode, or generate a pulse in pulse mode.
BUS
The instrument will be triggered if a 5AH command is sent via the RS-
232 interface.  See the Remote operation section for more details.
Voltage threshold operation
The DC Load can be set to only turn on if the voltage is at or above a set value.  Additionally, the 
load will turn off if the voltage drops below a second set value.  This feature works for any mode.  
An example of use would be to ensure an electronic system under test will not have power applied 
unless the supply voltage is above a certain value.  The DC Load would be put in series with the 
power supply.  Remote sensing would be turned on and the remote sensing connector on the back 
would be used to connect to the power supply's output voltage.
Use the menu items SYSTEM SET:VOLTAGE ON SET and  SYSTEM SET:VOLTAGE OFF SET to 
set the on and off voltage thresholds, respectively.
Example
 
    1:
     set the  VOLTAGE ON SET and  SYSTEM SET:VOLTAGE OFF SET both to 1.0 V. 
Set the instrument to constant current mode with the I-set key and set the current level to 0.1 A. 
Turn the load on by pressing the On/Off key.  When power is applied to the DC Load, the voltage 
must rise above 1.0 V before the load draws current from the source.  If the voltage drops below 1.0 
V on the load's terminals, the load will stop drawing current from the source.
Example 2:  Same as the previous example, except set the VOLTAGE ON SET to 1.0 V and the 
SYSTEM SET:VOLTAGE OFF SET 0.0 V.  As before, the load will turn on only when the voltage 
exceeds 1.0 V, but once it is "triggered" on, it will now stay on, even if the voltage drops to zero.
Passwords
If you enter the menu CONFIG:KEY LOCK SET, you can set a one to four digit password.  This 
password will be required when changing settings from the front panel.  The only operations allowed 
without entering the password are:
Shift + A
Shift
 + B
Shift
 + Short
Shift
 + Tran
Shift
 + Trigger
On/Off
To remove the password, enter the menu CONFIG:KEY LOCK SET and do not press any number 
keys, then press Enter.
Protection features
To protect external hardware when using the DC Load, you can set the maximum allowed values for 
current, voltage, and power.  These settings will override any settings subsequently made from the 
front panel using the I-setV-setP-set, or R-set keys.
8500 DC Load Series                    
Version:  September 2, 2009 
Page 40 of 76