Gw Instek PEL-3041 Electronic Load PEL-3041 Data Sheet

Product codes
PEL-3041
Page of 174
 
 
COMMNAND OVERVIEW 
87 
Query Syntax 
:RESistance:FREQuency? 
Related 
Commands 
:RESistance:DUTY 
Parameter  
<NRf> 
Sets the switching 
frequency in hertz. 
 
MINIMUM or MIN 
Minimum frequency 
 
MAXIMUM or MAX 
Maximum frequency 
Return parameter  <NR2> 
Returns the frequency in 
hertz. 
Example 
:RESistance:FREQuency 60 
Sets frequency to 60Hz. 
Query example 
:RESistance:FREQuency? 
>60 
Returns the switching frequency (60Hz). 
:RESistance:DUTY 
Set
Query
 
 
Description 
Sets the positive duty cycle for the CR dynamic 
mode switching frequency. This command along 
with the “:RESistance:FREQuency” command is 
used when the dynamic time setting is set to 
“Freq./Duty”. 
This command is only applicable for dynamic 
mode. 
Syntax 
:RESistance:DUTY {<NRf> | MINimum | MAXimum} 
Query Syntax 
:RESistance:DUTY? 
Related 
Commands 
:RESistance:FREQuency 
Parameter  
<NRf> 
Sets the duty as a 
percentage. 
 
MINIMUM or MIN 
Minimum duty 
 
MAXIMUM or MAX 
Maximum duty 
Return parameter  <NR2> 
Returns the duty as a 
percentage.