Gw Instek PEL-3211 Electronic Load PEL-3211 Data Sheet

Product codes
PEL-3211
Page of 119
 
 
PEL-3000 User Manual 
54
 
 
6.  Set the resistance or conductance level 
parameters using the scroll wheel and number 
pad.  
 
For Static mode, set CR A Value and/or CR B 
Value
 
For Dynamic mode, set Level1 and Level2.  
 
The maximum and minimum conductance/ 
resistance levels depend on the selected current 
range. 
7.  To add CV mode to CR mode (CR+CV), see 
8.  Set the remaining basic configuration settings 
such as the slew rate, and switching mode 
settings. See page 63 for details. 
Display 
Mode
CR
I Range
H 35A
V Range
L 15V
Function
Static
Configure
0.000
V
0.000
A
0.00
w
SlewRate
250.000
CR B Value
0.04082
CR A Value
0.04082
LOAD
RS232
A Value
Fine
03/Sep/2012
mA/uS
Ω
Ω
Mode
Current range
Voltage range
Active setting
Conductance/
Resistances
Conductance/
Resistance settings
 
 Note 
Basic CR mode configuration is complete. See 
page 63 for more configuration options. 
The current range and voltage range applies to all 