Gw Instek PEL-3211 Electronic Load PEL-3211 Data Sheet

Product codes
PEL-3211
Page of 119
 
 
PEL-3000 User Manual 
56
 
 
6.  Set the voltage level parameters using the scroll 
wheel and number pad.  
 
Set CV A Value and/or CV B Value. 
 
The maximum and minimum voltage levels 
depend on the selected voltage range. 
7.  Set the remaining basic configuration settings 
such as the response settings. See page 63 for 
details. 
Display 
Mode
CV
I Range
H 35A
V Range
L 15V
Response
Slow
Configure
0.000
V
0.000
A
0.00
w
CV B Value
15.000
CV A Value
15.000
LOAD
RS232
A Value
Fine
03/Sep/2012
V
V
Mode
Current range
Voltage range
Active setting
Current setting
Voltage settings
 
 Note 
Basic CV mode configuration is complete. See 
page 63 for more configuration options. 
The current range and voltage range applies to all 
the operating modes. 
CP Mode  
 
Description 
In Constant Power Mode, the unit will maintain 
a constant power by varying the current. For 
more details on CP mode, see the appendix on 
page 188.