OPERATION 
57 
 Warning 
If you change the mode or the range when the load 
is already on, the load will be turned off 
automatically.  
Operation 
1.  Make sure the load is off. 
2.  Press 
Main
3.  Select CP mode with the Mode[F1] soft-key.  
4.  Select the current range with the I Range[F2] 
soft-key.  
 
Range:  
High, Middle, Low 
 
5.  Select the voltage range with the V Range[F3] 
soft-key.  
 
Range:  
High, Low 
 
6.  Set the power level parameters using the scroll 
wheel and number pad.  
 
For Static mode, set CP A Value and/or CP B 
Value
 
For Dynamic mode, set Level1 and Level2.  
 
The maximum and minimum power levels 
depend on the selected current range. 
 
For static mode, the parameter that is set last 
becomes the ―active‖ setting. This will be shown 
in the Operation Status Panel. 
7.  To add CV mode to CP mode (CP+CV), see 
8.  Set the remaining basic configuration settings 
such as the slew rate, and timer settings. See 
page 63 for details.