Display 
2.596
V
500.06
mA
1.2982
w
CV B Value
15.000
CV A Value
15.000
LOAD
RS232
A Value
Fine
03/Sep/2012
V
V
OPP Alarm
Please Press Enter
To Clear Alarm
OPP
OCP indicator
Alarm message 
when OPP is set 
to Load Off
Mode
CV
I Range
L 0.7A
V Range
L 15V
Response
Fast
Configure
 
UVP 
 
Description 
If the UVP is tripped, the PEL-3000 will turn off 
the load. 
The UVP levels can be set from 0V to 10% 
higher than the rating voltage. 
Operation 
1.  Press 
Main
 > Configure[F5] > Protection[F1] 
and set the UVP Level.  
 
Range:  
UVP Level: OFF, 0~Rating voltage + 
10% 
Alarm 
 
The UVP indicator and a message will only 
appear on the screen when the input voltage is 
below the UVP level. The Enter key must be 
pressed to clear the alarm message.  
 
To clear the UVP indicator, remove the cause of 
the under voltage ~ i.e., increase the input 
voltage.