Gw Instek PEL-3211 Electronic Load PEL-3211 Data Sheet

Product codes
PEL-3211
Page of 119
 
OPERATION 
83 
Display 
2.596
V
500.06
mA
1.2982
w
CV B Value
15.000
CV A Value
15.000
LOAD
RS232
A Value
Fine
03/Sep/2012
V
V
UVP Alarm
Please Press Enter
To Clear Alarm
UVP
UVP indicator
Alarm message 
when UVP is 
tripped
Mode
CV
I Range
L 0.7A
V Range
L 15V
Response
Fast
Configure
 
OVP 
 
Description 
If the OVP is tripped, the PEL-3000 will turn off 
the load. 
The OVP levels can be set from 0V to 10% 
higher than the rating voltage. 
Operation 
1.  Press 
Main
 > Configure[F5] > Protection[F1] 
and set the OVP Level.  
 
Range:  
OVP Level: OFF, 0~Rating voltage + 
10% 
 
Note: To turn OVP off, set the OVP voltage greater 
than the current rating voltage + 10%. 
Alarm 
 
The OVP indicator and a message will only 
appear on the screen when the input voltage is 
below the UVP level. The Enter key must be 
pressed to clear the alarm message.  
 
To clear the OVP indicator, remove the cause of 
the over voltage ~ i.e., reduce the input voltage.