Automatic Trigger Options 
In the trigger menu, settings for automatic triggering can 
be changed as follows. (See also Chapter 1: “Displaying 
an Unknown Signal with Connect-and-View”
TRIGGER
 
Display the 
TRIGGER 
key labels. 
 
Note 
The 
TRIGGER 
key labels can differ depending on 
the latest trigger function used. 
F4
 
Open the 
TRIGGER OPTIONS 
menu. 
ENTER
 
Open the 
AUTOMATIC TRIGGER 
menu. 
If the frequency range of the automatic triggering is set to 
> 15 Hz, the Connect-and-View™ function responds more 
quickly. The response is quicker because the test tool is 
instructed not to analyze low frequency signal 
components. However, when you measure frequencies 
lower than 15 Hz, the test tool must be instructed to 
analyze low frequency components for automatic 
triggering: 
ENTER
Select >
 
1
 
H
Z
 and return to the 
measurement screen.