Triggering on Waveforms 
 
Triggering on Edges 
4
 
  
57 
Triggering on Edges 
If the signal is unstable or has a very low frequency, use 
edge triggering to obtain full manual trigger control. 
To trigger on rising edges of the input A waveform, do the 
following: 
TRIGGER
 
Display the 
TRIGGER 
key labels. 
 
F4
 
Open the 
TRIGGER OPTIONS 
menu. 
 
ENTER
 
Open the 
TRIGGER ON EDGE 
menu. 
 
 
When Free Run is selected, the test tool updates the 
screen even if there are no triggers. A trace always 
appears on the screen. 
When On Trigger is selected, the test tool needs a trigger 
to display a waveform. Use this mode if you want to 
update the screen only when valid triggers occur. 
When Single Shot is selected, the test tool waits for a 
trigger. After receiving a trigger, the waveform is displayed 
and the instrument is set to HOLD. 
In most cases it is advised to use the Free Run mode: 
ENTER
Select Free Run, jump to Trigger 
Filter
ENTER
Set Trigger Filter to Off
Observe that the key labels at the bottom of the screen 
have adapted to allow further selection of specific edge 
trigger settings: