Step and Ramp Operation
For step and ramp operation, the percent keys can be
used to set the milliamp output to 0 % of span, 100 % of
span, or step the output by 25 % of span.
used to set the milliamp output to 0 % of span, 100 % of
span, or step the output by 25 % of span.
Hands-free operation is possible with . Set the
Product to automatically and continuously step or ramp
the milliamp output from 0 % to 100 % and back.
Product to automatically and continuously step or ramp
the milliamp output from 0 % to 100 % and back.
The 100 % value is always 20 mA, but the 0 % value can
be 0 mA or 4 mA. This depends on how the mA span is
set. The 25 % step size is 5 mA or 4 mA accordingly.
be 0 mA or 4 mA. This depends on how the mA span is
set. The 25 % step size is 5 mA or 4 mA accordingly.
To use the manual step function:
1. Use the Main menu to set the Product to source or
simulate current.
2. Push
to set the output to 0 % of span.
3. Push
to set the output to 100 % of span.
4. Push
to step the output in 25 % of span
increments, from 0 % of span to 100 % of span and
back.
back.
Automatic Step and Ramp
To use the automatic step and automatic ramp functions:
1. Use the Main menu to set the Product to source or
simulate current.
2. The Product has separate auto ramp and step times.
Use the menu to set the ramp or step time.
3. Push
once to continually step the output from
0 % of span to 100 % of span and back in increments
of 25 % of span at the specified interval.
of 25 % of span at the specified interval.
4. Push
once more to go to auto ramp.
5. Push one of the percent keys, and twice, to turn
off the auto step and ramp.
6. Push
twice to continually ramp the output from
0 % of span to 100 % of span over the specified
interval, and then back over the specified interval.
interval, and then back over the specified interval.
7. Push one of the percent keys or once more, to
turn off the auto step and ramp.