Fluke Fluke-709 Calibrator, 4234350 Data Sheet
![Fluke](https://files.manualsbrain.com/attachments/9d57315a76f85bd9af23da46bf7ecaf7f248b87d/common/fit/150/50/7671c6748e9e70e4c5afb082a772388950496a1161e6cfdedf4b942e284b/brand_logo.png)
Product codes
4234350
Precision Loop Calibrator
Trim, Set, and Zero Menu
35
Use the selection knob to select and start the necessary
action. Push to go to the Function Select menu.
action. Push to go to the Function Select menu.
Trim 4 mA
If the operation mode is Communicator Only, this
function is not available and an error message is shown.
function is not available and an error message is shown.
A prompt to change the loop to MANUAL before you
proceed is shown. Push the selection knob to proceed.
Push to go to the Trim, Set and Zero menu.
proceed is shown. Push the selection knob to proceed.
Push to go to the Trim, Set and Zero menu.
While the HART device is changed to fixed output mode,
An error is shown if the HART device rejects the mode
change command. When the mode change is successful,
the screen in Figure 36 is shown.
An error is shown if the HART device rejects the mode
change command. When the mode change is successful,
the screen in Figure 36 is shown.
gzx45.eps
Figure 36. Trim 4 mA Screen
While the output to settles at 4 mA, the screen shows the
Product measurement. The measurement is updated
once a second.
Product measurement. The measurement is updated
once a second.
1. Push the selection knob to trim the HART device.
Stay on this screen to evaluate the result. An error is
shown if the HART device rejects the trim command.
shown if the HART device rejects the trim command.
2. Push
to change the HART device to normal
output mode and go to the Trim, Set, and Zero menu.
A warning to change the loop to AUTOMATIC is
shown first. An error is shown if the device rejects the
mode change command.
A warning to change the loop to AUTOMATIC is
shown first. An error is shown if the device rejects the
mode change command.