Precision Loop Calibrator 
 
Trim, Set, and Zero Menu 
  
37
 
Set Fixed mA Output 
If the operation mode is Communicator Only, this 
function is not available and an error message is shown. 
A warning to change the loop to MANUAL before you 
proceed is shown. Push the selection knob to proceed. 
Push  to go to the Trim, Set and Zero menu. 
As the HART device is changed to fixed output mode, an 
information screen is shown. An error is shown if the 
HART device rejects the mode change command. When 
the mode change is successful, the screen in Figure 38 is 
shown. 
 
gzx47.eps 
Figure 38. Set Fixed mA Output Screen 
 
This screen is used to set a fixed output and monitor the 
result with the Product measurement. The measurement 
is updated once a second. 
The range of values that can be set is 3.0 mA to 21.0 mA. 
1.  Use the selection knob to select and increment or 
decrement the value in steps shown by the selected 
decade. 
2.  Push and hold the selection knob to send the new 
value to the HART device and stay on this screen. 
An error is shown if the HART device rejects the 
value. 
3. Push 
 to change the HART device to normal 
output mode and go to the Trim, Set, and Zero menu. 
A warning to change the loop to AUTOMATIC is 
shown first. An error is shown if the device rejects the 
mode change command.