VXi VT1422A User Manual

Page of 529
Technology
VXI
Acquisition
Online at vxitech.com 
949 • 955 • 1VXI
214
Acquisition
Algorithmic Closed Loop Controller and 
Remote Channel Multifunction
VT1415A/1422A
VT1415A/1422A
To calculate accuracy over temperature range outside the 
±1 °C  range, results after *CAL are given by replacing each of 
the above error terms as follows: 
Replace 
 
(GainError)2 
with 
 
(GainError)2 + (GainTempco)2 
Replace 
 
(OffsetError)2 
with 
 
(OffsetError)2 + (OffsetTempco)2
Loop Control Specifications
 
Number of loops:
  
 
1 to 32
Default control 
   
 
PID
algorithm type:
Maximum VT1415A loop update rate 
for default PID algorithm:
 
(Note: VT1422A maximum sample rate is 25 kSamples/s, 
compared to 56 kSa/s for the VT1415A. The loop 
speeds of the VT1422A are reduced in same ratio.) 
 
 
1 loop:   
 
3 kHz
 
8 loops:   
 
1 kHz
 
32 loops: 
 
250 Hz
Custom algorithm development:
Language: 
 
 
Subset of C programming
language including if-
then-else, most math and 
comparison operations.
Variable types: 
 
 
Scalar local and global  
 
 
 
 
variables, array local and  
 
 
 
 
global variables.
Intrinsic functions: 
 
interrupt( ), writefifo( ),  
 
 
 
 
writecvt( ),writeboth( ), 
 
 
 
 
min( ), max( ), abs( ).
Other functions:   
 
Create own custom 
 
 
 
 
functions to handle  
 
 
 
 
transcendental operations.
I/O General
A total of eight (8) Signal Conditioning Plug-ons (SCPs) 
can be installed in most combinations of input or output 
configurations on a single VT1415A/VT1422A.
Power Available for SCPs
 
 
 
 
±24 V:   
1.0 A
 
 
 
 
5 V: 
 
3.5 A
General Specifications
 
VXI device type:
   
 
A16, slave only,    
 
 
 
 
Register based
Size:
 
 
 
 
C
Slots:
 
 
 
 
1
Connectors:
 
 
 
P1/2
Shared memory:
   
 
n/a
VXI buses:
 
  
 
TTL Trigger bus 
Drivers:   
 
 
VXIplug&play with
 
 
 
 
Source Code
Instrument Drivers - See the VXI Technology Website
www.vxitech.com for driver availability and 
downloading.
VT1415A
 