VXi Computer Hardware VX1410A User Manual

Page of 360
STATus Subsystem
3–52
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual
STATus:OPERation:VXITrigger:TRIGger?
Queries the time when the VXI TTL Trigger signal was asserted. The time is
relative to the mainframe power on time.
If there are multiple occurrences of the TTL Trigger signal, only the time of
the first assertion is saved.
When no Trigger is detected, the query returns a value of 0 in all fields. At power
on, or after starting the VXI monitor, the values of all fields are set to 0.
There is no command associated with this query.
STATus:OPERation:VXITrigger:TRIGger? <trigger>
Parameter
Query response
{TTL0trig ... TTL7trig}
<Hour>,<Minute>,<Second>
Not Applicable
None
None
Query:
STAT:OPER:VXIT:TRIG? TTL0
Response:
12,3,1
STATus:OPERation:VXITrigger:CONDition?
STATus:OPERation:VXITrigger:ENABle
STATus:OPERation:VXITrigger?
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands