IEEE 488.2 Common Commands
VX1410A & VX1420A IntelliFrame Mainframe Instruction Manual
3–165
*OPC
*OPC?
The command synchronizes command execution with the controller. The
command causes the device to set the OPC bit in the Standard Event Status register
after the previous command has completed execution.
The query causes an ASCII one to be placed in the output queue after the previous
command has completed execution.


None
Not Applicable
None
None
Command:

Query:

Response:

*WAI
Syntax
Parameters
Reset Value
Errors and Events
Dependencies
Examples
Related Commands