Measurement Specialties 9116 User Manual

Page of 143
Pressure Systems, Inc.                                                                                        Model 9116 User’s Manual 
 
 
 
Page 111 
 
www.PressureSystems.com 
 
Figure 5.5 
Solenoids in Module 
 
 
(4)  Ensure the three (3) solenoid manifold O-rings are present and free of contaminants.  
Replace the solenoid with the new one by carefully aligning and gently tightening the 
screws. 
 
(5)  Attach the wiring harness to the solenoid and connector P6 of the PC-322 Main Board 
Assembly. 
 
(6)  Reassemble the module.  
 
(7)  Test your scanner to enusre proper operation. 
 
 
5.1.6  
  Replacement of O-Rings 
 
Pressure Systems’ calibration valves include static and dynamic O-ring seals.  When used 
properly, the rated durability of the dynamic O-rings is in excess of 1,000,000 shifts of the 
calibration valve. 
 
The procedures described below should be used for replacement of all the O-rings in the Model 
9116
 Intelligent Pressure Scanner.  Use the tools and follow the general warnings already 
described at the start of Section 5.1
 
The material needed for the O-ring replacement can be acquired through the proper 
maintenance kit available from Pressure Systems.  Specifically needed for these procedures are 
calibration manifold and piston O-rings, Teflon cup seals, a fast evaporating cleaning fluid ( i.e. 
acetone, alcohol, Freon
®
, etc.), 50 psi dry air supply, and Krytox
®
 fluorinated grease (read 
product warnings and recommendations thoroughly).  Service of O-ring seals requires a 
clean working
 environment.  Introduction of contaminants to the O-ring or internal calibration 
manifold surfaces can result in internal pneumatic leaks. Inspection for contaminates generally 
requires some type of magnification device, such as a microscope
Solenoids