Pressure Systems, Inc.                                                                                        Model 9116 User’s Manual 
 
 
 
Page 41 
 
www.PressureSystems.com 
CALCULATE AND SET GAINS  (Command ‘Z’)
 
Purpose:    Instructs a module to calculate new gain coefficients, with either full-scale  
 
 
  pressure (or a specified pressure) applied to the specified channels.  These new  
 
 
  coefficients update part of the module’s internal calibration coefficient database,  
 
 
  used to convert any subsequent raw data (from any of the specified channels)  
 
 
  into engineering units data.  The new gain values are also returned in the  
 
 
  response. This command is sometimes called a Span or Span-only calibration. 
 
 
Command 
Zppppvv.vvvv]” 
 
‘Z’  is the command letter 
pppp’ is the  position field 
 vv.vvvv’ is an [optional] pressure value, preceded by a space character
.
 
 
Response 
g.gggg .. g.gggg” 
 
g.gggg’ are the actual  gain data values returned, each preceded by a 
space
 
Description: The  position field may have 0 or 4 characters.  If no  position field is specified,  
 
 
  gain coefficients for all module input channels will be calculated and returned.  If  
  
 
 
position field is specified, gain coefficients for only the channels whose bits are  
 
 
  set (=1) will be calculated and returned.  If the optional pressure value (vv.vvvv)  
 
 
  is specified, the  position field must be 4 characters, even when all channels are  
 
 
  to be specified. Gain values are returned in the response in order of highest  
 
 
  specified channel to lowest specified channel, with data formatted per an implied  
 
 
  decimal format (f=0). 
 
  Normally this command requires that the exact full scale input pressure be  
  applied to the affected channels.  The optional pressure value [ vv.vvvv] allows  
  the user to specify any suitable upscale pressure in the current engineering units.   
  For best results, pressures in excess of 90% of full scale should be applied.  A  
 leading 
space character must precede the pressure value parameter.  
 
  The desired calibrating pressure must be applied to all of the specified channels  
  and allowed to stabilize before this command is executed.  Such a pressure is  
  presumably generated by a separate model 903x calibrator module or suitable  
  user-supplied substitute.   
 
Notice that unlike the Calculate and Set Offsets (‘h’) command, this command 
does not automatically move a Model 9116 module’s calibration valve to its Cal 
position. A command to do this must precede this command. The reader is 
referred to Chapter 4, Section 4.3 for additional details concerning the 
performance of a Span Calibration.