Pressure Systems, Inc.                                                                                        Model 9116 User’s Manual 
 
 
 
Page 51 
 
www.PressureSystems.com 
 
Example: 
● 
Configure three (3) separate autonomous host delivery streams, and divide the module’s  
 
channels between them.  Channels (1-4) must be delivered to host as fast as possible,  
 
channels 5-8 may be delivered at half that rate, while the remaining channels 9-16 are  
 
delivered at half the previous rate.  All streams are generated continuously and  
 
synchronized with the internal clock at 100 msec., 200 msec., and 400 msec. periods,  
 
respectively.  Data are requested in single precision binary IEEE float format f=7). 
 
 
c 00 1 000F 1 100 7 0” 
 
c 00 2 00F0 1 200 7 0
 
c 00 3 FF00 1 400 7 0
 
 Read 
response: 
 
 
“A” 
 “A” 
 “A” 
 
To similarly acquire data at  “relative” rates (1, 2, and 4) using a periodic hardware 
trigger (assumed to also cycle at 10 Hz rate), enter the commands: 
 
 
c 00 1 000F 0 1 7 0” 
 
c 00 2 00F0 0 2 7 0” 
 
c 00 3 FF00 0 4 7 0” 
 
 Read 
responses: 
 
 
A” 
 
A” 
 
A” 
 
 
 
Note
 
The type of data delivered for each specified channel (when the 
streams are started) is EU pressure unless sub-command “05” is 
also executed to select
 other types of data in each stream.