Hitachi Microscope & Magnifier S-3400N User Manual

Page of 466
3.2.4 
3 - 11 
(2)  Setting the Specimen Height and Specimen Height and Size 
The specimen height and the specimen size measured in (1) are set.    This setting limits the 
amount of the movement of the stage (Z and T direction), and protects the specimen and 
and components inside the specimen chamber from damage by contact of both.   
 
(a)  Moving the Stage to the Specimen Exchange Position 
The specimen size and height can be set only at the exchange position. 
If the Set button is disabled, (grayed out), click the Exchange Z65 [HOME Z65] button in 
Fig. 3.2-10 (a).    The stage will move to Exchange position. 
 
(b)  Setting the Specimen Height and Specimen Size 
On the SPECIMEN block on the Stage tab, click the Set button in Fig. 3.2-10 (b). Set 
Sample Size / Detector window will open.   
 
 
(a) Exchange Z65 button 
(b) SPECIMEN block 
 
 
 
 
 
 
 
Fig. 3.2-10    Exchange Z65 Button and SPECIMEN Block