Hitachi Microscope & Magnifier S-3400N User Manual

Page of 466
3.9.14 
3 - 145 
3.9.14  Optional Setting 
 
Operation environments are set on the Optional Setup dialog 
window.  
To open the Optional Setup dialog window, select the Optional 
Setup command from the Option menu or click the Opt Setup 
button 
  on the tool button area of the Operation panel. 
The dialog window has 6 tabs.    Each has functions as follows. 
 
NOTICE:  Setting on each tab is applied when the OK button is 
clicked. 
 
(1)  General tab: general setting 
 
(a)  RASTER ROTATION area 
WD-Rotation Link selection specifies if the image 
rotation caused by the objective lens magnetic field will 
be compensated automatically or not.   
For normal operation, it is strongly recommended to 
enable the automatic compensation function. 
 
(b)  MICRON MARKER area 
When Sizable is selected, the length of the micron 
marker will be adjusted so the length indication does 
not have a fraction. 
If Fixed is selected, the length is fixed at 30 mm on 4 
× 
5 inch photograph.    In this case, the length value 
indication some times has a fraction, especially by coarse focusing. 
 
(c)  Magnification fraction adjustment 
Put check mark to this checkbox to enable the auto magnification fraction adjustment. 
Magnification value will have fraction by focusing operation.   
The auto fraction adjustment calculates and set nearest even magnification value when 
starting image capturing or direct photographing. 
 
(d)  DATA ENTRY/MEASUREMENT 
This box allows you to specify the character font to be used in the Data Entry and 
Measurement functions.  
Specify the font in Font Name, and the character size in Font Size. 
 
NOTICE: In Font Name, fonts that have an @ symbol at the beginning represent horizontally 
oriented fonts. 
 
 
Fig. 3.9-39  General Tab