Hitachi Microscope & Magnifier S-3400N User Manual

Page of 466
 
APPENDIX - 1 
APPENDIX 
Message List 
 
The table below lists messages that can appear when the S-3400N is being used.   
Messages without a message code are for instructions, simple notifications, or confirmation 
messages; they are not included in the table below.   
Message codes listed below that are marked with an * appear as error codes. 
 
Code 
Message Text 
Explanation and Countermeasure 
*1301 Repair 
unsuccessful. 
*1302 Optimize 
unsuccessful. 
Failed to repair or optimize database file using SEM Data 
Manager - Tool – Database Utility or S-4800 – Help – 
Maintenance function.    If you tried and failed to repair 
database when the code 2014 error appeared, contact a 
service engineer. 
2000  Invalid input data 
2001  Invalid data (out of range) 
2002  Only a number is acceptable. 
2003 Invalid 
Data. 
These messages will be shown when input data is not 
correct.    Confirm acceptable data and then, input again. 
 
2004  The name is used in the system. S-3400N uses this login name.    Use other names. 
2007  Invalid login name. 
Use a login name already registered.    [S-3400] is the 
default login name. 
Note that capital and small letters are distinguished for login 
names.  Confirm your input. 
2008  The password you typed is 
incorrect. 
You entered incorrect password.    Note that capital and 
small letters are distinguished for login names.     
Confirm your input. 
*2010  Specified file is different from the 
condition file for PC_SEM.    SEM 
Parameters will be initialized. 
The file for operation condition is not correct.    When this 
message is shown, check if other application is using files 
of extension pm1. 
*2014  Database not found 
Failed to access database files of S-4800 or SEM Data 
Manager.    If the message is shown when you are using 
SEM Data Manager, try to repair the file [SDM.sdb] using 
Tool – Database Utility.    If it is shown when you are 
operating SEM, try to repair files using   
Help - Maintenance function. 
2016  Specified file not moved because 
source and destination folders 
are the same. 
This message will be shown when you have specified the 
same directory as of the source files for the target directory 
in Batch Process - Move File command.    Specify a 
directory other than that of source files for the target 
directory. 
2017  Is already Exist. 
This message will be shown when the input User name (or 
login name) is already used.    Specify another name. 
2031  Use as Magnification Lower than 
x5,000 
This message will be shown when you click the Get Image 
button at a higher magnification than 
×5,000.