1.2.5  Stray Magnetic Field 
 
Image disturbance will be negligible when the stray magnetic field conditions shown in Tables 
1.2-1 are satisfied at the instrument installation place.    Before proceeding to installation, check 
the stray magnetic field conditions as required.    Make sure that the conditions shown in the 
tables are satisfied.   
If there is a large-sized magnet clutch or power cable for other equipment in the vicinity of this 
instrument, abrupt variation in current or magnetic field may take place.    Avoid installing the 
instrument in such a location.   
If the instrument is installed in an environment having intense stray magnetic fields, the scanning 
electron beam is forced to tremble due to the magnetic field, causing abnormal expansion, 
shrinkage, distortion, unintentional shift, or whisker noise during imaging.    These adverse 
effects occur differently depending upon whether the frequency component of stray magnetic 
field is synchronous with the power current frequency or not.    Therefore, the allowable value of 
stray magnetic field component synchronous with power current frequency is different from that 
of stray magnetic field component asynchronous with power current frequency (DC magnetic 
field variation). 
•  The stray magnetic field component synchronous with power frequency appears as image 
distortion or deformation on the CRT display (Fig. 1.2-1).    This is because the electron 
beam is scanned in synchronization with the power frequency.     
The maximum allowable value of stray magnetic field component synchronous with power 
frequency is specified so that no adverse effect will be given to CD measurement 
reproducibility.  
•  The stray magnetic field component asynchronous with power frequency affects imaging 
directly.    That is, movement of the electron beam due to variation in external magnetic field 
appears as unintentional shift or whisker noise in imaging (Fig. 1.2-2).     
Even a slight variation in external magnetic field may result in conspicuous irregularity of the 
image.    The maximum allowable value of stray magnetic field component asynchronous 
with power frequency is specified so that CD measurement reproducibility will not be 
affected by image shift or whisker noise.   
The effect on the scanning electron beam varies also according to whether the stray magnetic 
field component is in the horizontal or vertical direction, and the allowable value differs for each 
of these components.    Given below are the allowable values stipulated under the resolution 
guaranteeing conditions (high resolution observation conditions), and under conditions different 
from these. Image disturbance may occur even if the stray magnetic field at the site satisfies the 
allowable values given in Tables 1.2-1.     
In general, the effect of a stray magnetic field (wavering of scanning electron beam) is inversely 
proportional to the square root of the accelerating voltage, provided the other parameters are 
kept constant.    The effect increases nearly proportionally to the working distance.