Hitachi Microscope & Magnifier S-3400N User Manual

Page of 466
2.1    Control Knobs and Switches on Main Unit 
2 - 4 
2.1.3    Specimen Stage (Type I) 
 
Figure 2.1-3 shows the Type I specimen stage. 
 
 
 
Fig. 2.1-3    Specimen Stage (Type I) 
 
•  [X] control:  Moves specimen in X direction within a range from 0 to 80 mm. 
•  [Y] control:  Moves specimen in Y direction within a range from 0 to 40 mm. 
•  [Z] control:  Moves specimen in Z direction or shifts WD (Working Distance) within a range 
from 5 to 35 mm.    (X-ray analysis position is at 15 mm working distance) 
•  [R] control:  Rotates specimen through 360° (continuously variable) 
•  [T] control:  Tilts specimen within a range from -20° to 90°. 
•  [Touch sensor and absorbed electron measurement terminal]:  This terminal is used for the 
measurement of the irradiation 
current of electrons on the 
specimen (normally it is used 
as a touch sensor). 
Touch Sensor 
:  This observes whether the specimen or the stage come in contact 
with components inside the specimen chamber 
The sensor will inform with the buzzer and the message when 
coming in contact. 
Absorbed Current Terminal :  The absorbed current can be measured by removing the cable for 
the touch sensor, and connecting a minute ammeter instead. 
[R] control 
[Z] control 
[Y] control 
[T] control
 
Touch Sensor and 
Absorbed Current 
Terminal 
[X] control