Standard Operations
To add a fixed IP address to the computer:
1. In the Windows Start menu click Programs, Settings, Control Panel, Network 
Connections.
2. Right-click the Local Area Connection (LAN) option and select the Properties 
command.
3. In the LAN Connection dialog box select Internet Protocol (TCP/IP) and press the 
Properties button.
4. In the General tab select Use the following IP address radio box.
5. In the IP address box enter the computer IP address. 
6. In the Subnet mask box enter 255.255.255.0.
7. Confirm all settings. 
To run the LAN connection and start the web browser:
1. On the sc1000 controller switch to the measured value display.
2. Connect the computer to the service port on the sc1000 display module. Use the 
standard Ethernet RJ45 cross-over interface cable (LZX998).
3. Start the web browser.
4. Enter the IP address of the sc1000 controller (by default: 192.168.154.30) in the web 
browser address box.
5. The sc1000 log on screen is displayed.
6. Enter the password. The password is set in the sc1000 controller software under 
SYSTEM SETUP, BROWSER ACCESS, LOGIN PASSWORD.
7. The sc1000 controller can be administered remotely.
5.13.3 Set up the dial-up connection 
To set up a dial-up connection between a computer and the sc1000 controller certain 
settings are necessary.
To set the sc1000 controller settings:
1. Connect the external GSM antenna to the display module (refer to 
).
2. Insert the SIM card into the display module (refer to 
). 
3. Enter the PIN in SYSTEM SETUP, GSM MODULE, PIN.
4. Confirm with 
ENTER
.
5. Select SYSTEM SETUP, GSM MODULE, EXTERNAL DIAL-UP, ALLOW.
6. Confirm with 
ENTER
.
7. Enter a password for browser access in SYSTEM SETUP, BROWSER ACCESS, 
LOGIN PASSWORD.
8. Confirm with 
ENTER
.