Emerson HASXEE-IM-HS User Manual

Page of 502
Emerson Process Management GmbH & Co. OHG
 8-39
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
06/2010
8
Troubleshooting
8.4   Troubleshooting on Components
R60
Test connector 
P4
P4.15
  
P4.16
+
_
Step 3:
Locate test connector P4 to 
measure 
the bridge voltage:
P4.16 Bridge voltage (+)
P4.15 Bridge voltage (-); GND
CAUTION! 
Do not short-circuit pins!
Alternatively the GND signal (-) is ac-
cessible on the main board BKS, too: 
Locate X11 (
 fig. 8-3, page 8-16) .
The bridge voltage depends on range 
and sample gas and should be bet-
ween 3V and 5V.
Only if the WAP 100 board has been 
replaced, it is necessary to adjust the 
voltage with potentio-meter R60.