Fluke Digital Camera 1526 User Manual

Page of 41
Fluke  1520
Users Manual
32
Test Voltages
250 V, 500 V, 1000 V
Accuracy
+20 %, -0 %
Nominal Current
1 mA
Number of measurements
5,000
Circuitry Protection
test inhibited if 
 30 V ac or dc at inputs
Lo 
Ω
Display Range
40.00 
Ω
Measurement Range
0.10 
Ω
 to 40.00 
Ω
Accuracy
2 % + 2 counts
Resolution
0.01 
Ω
Analog Bar Graph
0 to 100 
Ω
Open Circuit Voltage
6 V typical
Short Circuit Current
200 mA minimum, (0.2 to 2.0 
Ω
 for >20%
battery capacity remaining).
Test Leads Zero
Ω
Number of measurements
5,000
Circuitry Protection
test inhibited if 
 30 V ac or dc at inputs