Fisher & Paykel MR730 User Manual

Page of 64
57
Ref. 185040646 MR730/720/700/480 Technical Manual ¯ Revision F ¯ Issued March 2001
Connect the Heater Wire cable to the humidifier.
If a gas supply is used, set to 5- 10 l/minute, if no gas supply used, the breathing tube should be supported
from the patient end so that it drops vertically over its entire length to the top of the chamber.
E.2.3
D
YNAMIC 
T
EST 
S
EQUENCE 
(MR730/720/700)
1
Turn the humidifier on, ensure the heater wire LED is on for a MR720 or MR730.
2
Adjust set temperature to 36.0 ± 0.1 °C and Chamber Control to 0.
3
NOTE: At low ambient temperatures (below 22 °C) a low temperature alarm may occur, if this occurs, press mute
and restart the test , the breathing circuit will eventually warm up and the humidifier will start controlling.
4
Check after 30 minutes:
Digital display is 36.0 ± 0.3 °C
Chamber outlet temperature is 36.0 ± 2.0 °C
If the unit is alarming or has not reached 36 ± 0.3 °C, press the mute button if necessary and leave for a further
15 minutes.
5
Then check the following in ‘Engineering Mode’ (see appendix A2):
set
36 ± 0.1 °C
offset
0
Edc
 0
Cdc
 0
hp
 36 °C
If unit passes this test it is controlling correctly and can be put back in use.
E.3
P
ERFORMANCE 
C
HECK 
(MR480)
E.3.1
E
QUIPMENT 
S
ETUP
Set up a non heated breathing circuit, see typical setup below:
Setup Notes:
Patient delivery tubing can be single use or reusable.
A gas supply of 2 – 10 l/minute to the chamber is recommended, if no gas supply is used, then the breathing circuit
must be 22mm diameter to ensure adequate airflow.
Use a dual airway temperature sensor, such as 900MR560.
  Use a Fisher and Paykel  humidification chamber, such as a MR200 series, filled with cold water to maximum level (not
a MR290).