Fisher & Paykel MR730 User Manual

Page of 64
58
Ref. 185040646 MR730/720/700/480 Technical Manual ¯ Revision F ¯ Issued March 2001
E.3.2
S
ETUP
Fill the humidification chamber with water and slide on to the heaterbase.
If a gas supply is used, set to 2- 10 l/minute, if no gas supply used, the breathing tube should be supported
from the patient end so that it drops vertically over its entire length to the top of the chamber.
3
Insert the middle sensor of the temperature probe into the connector on top of the chamber , and the end
sensor of the temperature probe into the connector at the end of the breathing tube, plug the temperature probe
plug in to the humidifier.
E.3.3
D
YNAMIC 
T
EST 
P
ROCEDURE 
(MR480)
1
Turn the humidifier on.
2
Turn Temperature Set control to 31.0 ± 0.1 °C, check that correct setting flashes on display after adjustment.
The airway circuit under the low flow conditions of the test will warm up only slowly and may give a low
temperature alarm. If this is muted the circuit will eventually warm up and start controlling.
3
Check after one hour that digital display is 31.0 ± 1.0 °C.
If temperature is still outside these limits leave the test going for a further 15 minutes.
4
If displayed temperature is within 31.0 ± 1.0 °C, press chamber temp button and check chamber temperature is
35 to 66 °C.
5
In Engineering menu (appendix A2) check:
hp (heaterplate temperature) > 36 °C
Set = 31.0 ± 0.1°C
6
Check that disconnecting the temperature probe gives an audio alarm and Temp Probe LED flashes.
7
Press Mute button and check the alarm is silenced.
8
If the Humidifier passes these tests it is controlling correctly and can be returned to use.