SCPI Commands – an Introduction 
62 
Series N5700 User’s Guide  
Parameter Types 
Data programmed or queried from the power supply is ASCII. The 
data may be numerical or character string. 
Numeric Parameters 
Symbol Response 
Formats 
<NR1> 
Digits with an implied decimal point assumed at the right of the 
least-significant digit. Examples: 273 
<NR2> 
Digits with an explicit decimal point. Example: .0273 
<NR3> 
Digits with an explicit decimal point and an exponent. Example: 
2.73E+2 
 
Parameter Formats 
<NRf> 
Extended format that includes <NR1>, <NR2> and <NR3>. 
Examples: 273     273.   2.73E2 
<NRf+> 
Expanded decimal format that includes <NRf> and MIN     
MAX. Examples: 273    273.     2.73E2     MAX.     
MIN and MAX are the minimum and maximum limit values that 
are implicit in the range specification for the parameter. 
<Bool> 
Boolean Data. Example: 0 | 1 or ON | OFF 
Discrete and String Parameters 
Symbol Response 
Formats 
None 
Discrete parameters are used to program settings that have a limited 
number of values (like RST, RCL0, BUS). You can mix upper- and 
lower-case letters. Query responses will always return the short form 
in all upper-case letters
 
“string” 
String parameters can contain virtually any set of ASCII characters. A 
string must begin and end with matching quotes; either with a single 
quote or a double quote. You can include the quote delimiter as part of 
the string by typing it twice without any characters in between.
 
Suffixes and Multipliers 
Class 
Suffix 
Unit 
Unit with Multiplier 
Current A 
ampere  MA 
(milliampere) 
Amplitude V 
volt 
MV 
(millivolt) 
Time S 
second MS 
(millisecond) 
Common Multipliers 
1E3 K  kilo   
1E-3 M  milli   
1E-6 U  micro