Status Commands 
72 
Series N5700 User’s Guide  
The Standard Event group is programmed with Common commands 
as described later in this section. Common commands also control 
additional status functions such as the Service Request Enable and 
the Status Byte registers. 
STATus:PRESet 
This command sets all defined bits in the Operation and Questionable 
PTR registers. The command clears all defined bits in the Operation 
and Questionable NTR and Enable registers. 
STATus:OPERation[:EVENt]? 
This query returns the value of the Operation Event register. The 
Event register is a read-only register, which stores (latches) all events 
that are passed by the Operation NTR and/or PTR filter. Reading the 
Operation Event register clears it. The bit configuration of the 
Operation status registers is as follows: 
 
Bit Position 
15-11 
10 9 8 7-6 
5 4-0 
Bit Value 
− 
1024 
− 
256 
− 
32 
− 
Bit Name 
− 
CC 
− 
CV 
− 
WTG 
− 
CC  = The output is in constant current  
CV  = The output is in constant voltage 
WTG  = The unit is waiting for a transient trigger 
STATus:OPERation:CONDition? 
This query returns the value of the Operation Condition register. 
That is a read-only register, which holds the live (unlatched) 
operational status of the power supply. 
STATus:OPERation:ENABle <NRf> 
STATus:OPERation:ENABle?  
This command and its query set and read the value of the 
Operational Enable register. This register is a mask for enabling 
specific bits from the Operation Event register to set the operation 
summary bit (OPER) of the Status Byte register. This bit (bit 7) is the 
logical OR of all the Operational Event register bits that are enabled 
by the Status Operation Enable register. The Preset value = 0.