Verification 
94 
Series N5700 User’s Guide  
Constant Voltage Tests 
NOTE 
Refer to the appropriate test record form for the instrument settings of the 
model you are checking. 
Voltage Programming and Readback Accuracy 
Test category = performance, calibration 
This test verifies that the voltage programming and measurement 
functions are within specifications.  
1  Turn off the power supply and connect a DVM directly across the 
+S and -S terminals as shown in figure A. Do not connect a load. 
2  Turn on the power supply and program the output voltage to zero 
and the output current to its maximum programmable value 
(Imax) with the load off. The CV annunciator should be on and 
the output current reading should be approximately zero. 
3  Record the output voltage readings on the digital voltmeter 
(DVM) and the front panel display. The readings should be within 
the limits specified in the test record card for the appropriate 
model under Voltage Programming and Readback @ 0 V.  
4  Program the output voltage to its full-scale rating. 
5  Record the output voltage readings on the DVM and the front 
panel display. The readings should be within the limits specified 
in the test record card for the appropriate model under Voltage 
Programming and Readback @ Full Scale. 
CV Load Effect 
Test category = performance 
This test measures the change in output voltage resulting from a 
change in output current from full load to no load. 
1  Turn off the power supply and connect a DVM and an electronic 
load as shown in figure A. 
2  Turn on the power supply and program the output current to its 
maximum programmable value (Imax) and the output voltage to 
its full-scale value. 
3  Set the electronic load for the output’s full-scale current. The CV 
annunciator on the front panel must be on. If it is not, adjust the 
load so that the output current drops slightly. 
4  Record the output voltage reading from the DVM. 
5  Open the load and record the voltage reading from the DVM 
again. The difference between the DVM readings in steps 4 and 5 
is the load effect, which should not exceed the value listed in the  
test record card for the appropriate model under CV Load Effect.