RayTek MI Miniature Infrared Sensor User Manual

Page of 95
Programming Guide
 
76 
MI 
11.7 Multiple Units (RS485 Multidrop Mode) 
Up to 32 units can be connected within a RS485 network, see section 
5.6 
  on  page  31.  To  direct  a 
command  to  one  unit  among  the  32  possible,  it  is  necessary  to 
„address“  a  command.  Therefore,  a  3‐digit  number  is  set  prior  the 
command.  The  3‐digit  number  is  determined  between  001  and  032. 
Exception: Broadcast command. 
If a command is transferred, starting with the 3‐digit number 000, all 
units  (with  addresses  from  001  to  032)  connected  will  get  this 
command – without to send an answer. 
Note:  A  unit  with  the  address  000  is  a  single  unit  and  not  in 
multidrop mode. 
XA=024  will set address to 24 (unit must not be in multidrop mode) 
Changing an address: 
(e.g. the address is change from 24 to 17) 
command 
answer 
„017?E“ 
„017E0.950“ 
„017XA=024“ 
“017XA024” setting of a new address 
„024?E“ 
„024E0.950“ 
Example for the broadcast command: 
command 
answer 
“024?E” 
“024E0.950” 
“000E=0.5” 
will be executed from all units, no answer 
“024?E” 
“024E0.500” 
“012?E” 
“012E0.500”